Handheld Particle Counters – Models 3888 and 3889

Description

Handheld particle counters for accurate measuring and monitoring

The Model 3888 (3-channel) and 3889 (6-channel) Handheld Particle Counters utilize the same proven components that made the original handheld particle counters highly reliable.

These instruments meet or exceed the latest ISO 21501-4 standards and are fully RoHS compliant.

The devices are equipped with a base/stand that enables advanced communication capabilities such as Wi-Fi, Ethernet, and RS485, along with a larger full-color touchscreen display. They also allow real-time monitoring when connected to a PC.

Key Advantages

  • Model 3888 measures particle sizes: 0.3 µm, 0.5 µm, 5.0 µm
  • Model 3889 measures particle sizes: 0.3 µm, 0.5 µm, 1.0 µm, 3.0 µm, 5.0 µm, 10.0 µm
  • Easy-to-use interface displayed on a 4.3-inch color LCD resistive touchscreen
  • Onboard data storage up to 10,000 measurements
  • Generates Microsoft Excel-compatible .CSV files without additional software
  • Data can be downloaded via USB memory
  • Kanomax Measuring Software provides real-time data display and graphing when connected to a PC
  • Multi-language support: English, Spanish, Japanese, and Chinese

Applications

  • Monitor and verify cleanrooms (pharmaceutical, semiconductor, food industry)
  • Monitor surgical rooms
  • Filter testing
  • IAQ (Indoor Air Quality) investigation
  • Data center cleaning

Technical Specifications

)

Main Unit Specifications
Product Handheld Particle Counter
Model 3888 3889
Particle Measurement Sizes 3 channels (0.3, 0.5, 5.0μm) 6 channels (0.3, 0.5, 1.0, 3.0, 5.0, 10.0μm)
Flow Rate 0.1CFM (2.83L/min)

Accuracy: ±5% (Compliant with JIS B9921 and ISO21501-4)

Maximum Measurable Concentration 10% for 2,000,000 particles/CF

(compliant with JIS B9921 and ISO21501-4)

Counting Efficiency 50±20% (for PSL particles near the minimum measurable size)

100±10% (for PSL particles of 1.5 to 2 times as large as the minimum
measurable size)

(Compliant with JIS B9921 and ISO21501-4)

False Count ≦1 particle/5 minutes

(Compliant with JIS B9921 and ISO21501-4)

Size Resolution ≦15% (for PSL particles near 0.3μm)

(Compliant with JIS B9921 and ISO21501-4)

Display 4.3 inch color LCD, Resistive touch panel
Communication Standard USB (Host: for printer and USB flash memory, Device: for PC)
Stand (option) RS485, Ethernet, and WiFi connection points in the base/stand
Recording Media Media Internal memory
Number and Format Up to 10,000 records in CSV format
Language English, Japanese, Chinese, and Spanish
Power Internal Supply Li-ion rechargeable battery (Replaceable)
External Supply AC adapter – Input 100 to 240 V
Continuous Operation Time 5 hours (Single measurement)
Operating
Environment
Main Body From 10 to 40 ℃, from 0 to 85%RH (With no condensation)
Probe From 0 to 50 ℃, from 2 to 98%RH (With no condensation)
Dimension 213 × 100 × 55 mm
Weight 650g
Hydrothermal
Measurement
(Available as an
option for the 6-
channel portable
particle counter
Measurement Range (Temperature) From 0 to 50℃

(Humidity) From 2.0 to 98.0%RH

Display Resolution (Temperature) 0.1℃

(Humidity) 0.1%

Humidity Accuracy ±3.0% (30 to 85%RH), ±5% (other range of humidity)
Temperature Accuracy ±0.5℃
Response Time Approximately 60 seconds or less (90% response)